Darlisha
Laura Ball posted a reply on Re: Telcor/QML Super users?
Tuesday, Oct 25, 2022 at 8:09 AM
Said: I currently have i-STAT, Clinitek, Nova glucose and GeneXpert Express instruments set up in Telcor. With Telcor I can manage users and see error messages from failed tests (ie: flow errors on the Nova glucose instrument). I ...
Brandy Kenny posted a reply on Re: Telcor/QML Super users?
Tuesday, Oct 25, 2022 at 8:06 AM
Said: Not a super user, but I can tell you my clinitek are connected to my QML. I don't have Freestyle but I am in the process of getting a quote to get iSTAT and have them connected.
Tony Dempsey posted a new Topic
Monday, Oct 24, 2022 at 7:28 PM
Telcor/QML Super users?
Hello,
I am looking for a Telcor/QML super user who wouldn't mind connecting with me so I could pick their brain. I am wanting to connect all my devices to QML to then send results to Epic and ditch uniPOC. Freestyle, clinitek, istat are the main devices I am thinking about.
I think it is possible to manager my users, devices, and monitor patient/qc results all in QML, I just need confirmation and some guidance.
Thanks
antony.dempsey@straub.n...
I am looking for a Telcor/QML super user who wouldn't mind connecting with me so I could pick their brain. I am wanting to connect all my devices to QML to then send results to Epic and ditch uniPOC. Freestyle, clinitek, istat are the main devices I am thinking about.
I think it is possible to manager my users, devices, and monitor patient/qc results all in QML, I just need confirmation and some guidance.
Thanks
antony.dempsey@straub.n...
Danielle Ennis posted a new Topic
Monday, Oct 24, 2022 at 6:54 PM
Validation of high pCO2 (>105) for ABL90
Hi everyone,
We are in the process of validating new ABL90s. The the highest pCO2 I could get using Radiometer-supplied linearity was 105 mmHg. Our RT department uses RapidPoints and are able to get up to 170 mmHg with their Siemens linearity. Has anyone been able to validate a higher pCO2 on ABLs? What are you using?
Thanks!
Danielle
We are in the process of validating new ABL90s. The the highest pCO2 I could get using Radiometer-supplied linearity was 105 mmHg. Our RT department uses RapidPoints and are able to get up to 170 mmHg with their Siemens linearity. Has anyone been able to validate a higher pCO2 on ABLs? What are you using?
Thanks!
Danielle
MOST ACTIVE MEMBERS
SHARED LINKS
The Latest Point-of Care Technologies: Preparing for an Era of Change
Shared byPeggy Mann onJul 7, 2020 9:37 PM
AACC’s International CPOCT Online brings you the latest on new and ...
Shared byPeggy Mann onJul 7, 2020 9:37 PM
POC Boot Camp in Pittsburgh, PA
Shared byPeggy Mann onJul 9, 2019 9:05 AM
Interactive format. See the agenda please for topics (they are cons...
Shared byPeggy Mann onJul 9, 2019 9:05 AM
AACC POC Bootcamp and Beyond (East Coast)
Shared byPeggy Mann onSep 29, 2017 8:01 AM
Communicating, connecting, and collaborating! Flyer available at th...
Shared byPeggy Mann onSep 29, 2017 8:01 AM
Southwest Regional POC Group Annual Meeting_Dallas, Texas
Shared byPeggy Mann onSep 29, 2017 7:57 AM
All are welcome! This year the annual meeting takes a different for...
Shared byPeggy Mann onSep 29, 2017 7:57 AM
UPCOMING EVENTS
This area will display upcoming events as they approach on the calendar
Add New Module