Said: We run the 4 levels (L1, L3, HCT-a and HCT-b) on new shipments in duplicate on only one device. First of the month we run the four levels on all lots we have in the system, so some months its one lot (4 cards) and some months we have 2 ...
Said: We try to order once a month, we run L1, L3, HCT-A, HCT-B and one patient sample on old and new lot.  We have 120+ devices so we run one device from each CLIA/CAP location.
Said: Our lab also is interested in doing a lab developed test for using the iSTAT for pH for pleural fluid.  Have you gotten any insight as to how to go about doing this? Our current pH meter needs to be calibrated every day, which is a...
Said: For new lots, I run 4 cards. L1 and L3 and HCT-A and HCT-B. This is what our implementation specialist requested for us to order. For 30 day QC, we run the same 4 levels on each device. We have 7 devices. (I think this is overkill becau...