Said: For 6 months you dont have to do Accuracy and Precision. For instrument to Instrument there is no required number. Most people do 5 because EP evaluator requires a minimum of 5 points for calculation. I have inspected lab that ...
Said: If just Bi-annual and not new setup/install then... ·         Method to Method: Study between the iSTAT and the Roche Cobas. 1.  min of 5 patient correct. ·    &n...
Said: I am not an iSTAT user, but what you described above sound like a new instrument verification.  For semi-annual verification, you don't need accuracy or precision.  For iSTAT to iSTAT, refer to your institution's policy.&n...