Said: Good Morning!We don't use that product from CAP, but if my team was having issues with calibration verification and the issue was with the pO2 and an external sample (as yours is), I would review whether it was too high or too low a...
Said: Hi Kim-What criteria are you using for precision where your HCT would fail?To specifically answer your question, if we have already validated the cartridges already being used on our iSTAT fleet, the new/replacement iSTAT, and a...
Said: We only run Creatinines on ours, but when we've gotten a new/replacement device we run the cal ver (levels 1-5) with 2 runs/each level and plot as a linearity. And then run a precision using 10 runs of both levels of QC. I think EP eval...