Said: @ Patricia Raucholz.  Do you then have reference ranges for the corrected gas reported?  
Said: We do correct for temp on arterial blood gases on our GEM 5000. We only report the corrected gases, not the values at 98.6.I am not sure how we validated this since we have been doing it for years. CLSI BloodGasC46A2 document may be...
Said: We also allow all results to cross. The only exception is <10 mg/dL - those I confirm it wasn't an error, there is a repeat immediately and/or confirmatory lab draw sent. If it meets those criteria I send it across to the LIS. <...
Said: We also let all results cross. Performing staff has been instructed to add a comment to results they may question and to repeat testing stating it is a repeat. We monitor results sent with comments and all critical result in the Roche I...