Said: Yes, we have iSTAT set in DE for the operator to select the sample type. In DE there can only be one reference range per analyte. The specific sample type reference ranges and if applicable critical ranges are in Epic.
Said: Does anyone have the ISTAT set to where the operator selects the sample type?Do you have two separate reference ranges?
Said: For several reasons, number 1 being if the sample is collected into a green top - then ambient air inside the green top tube will vastly change the PO2 especially when sent through a pneumatic tube station. For most specimens, if collec...
Said: Debra Stark, the best place to look is a recent CAP or API survey. They have calculated the SD for you for each analyte for the 5 specimens in the peer group. I look for one with a value close to the one I used for my method comp, then ...