Siemens Multistix 10 SG

1 followers
0 Likes

Anyone seeing an increase in false positive (Trace-Intact) blood results using Multistix 10 SG lot number 704054? We use them on the Clinitek Status+ and have seen the number of failures on our negative QC quadruple and have also had an unacceptable proficiency result.

1 Reply

We do not currently have that lot number. I have noticed however, depending on the control manufacturer, Negative-Trace is an acceptable range on Blood for Level 1/Negative.

Reply
Subgroup Membership is required to post Replies
Join POCT Listserv now
Tom Wilson
over 7 years ago
1
Reply
0
Likes
1
Followers
551
Views
Liked By:
Suggested Posts
TopicRepliesLikesViewsParticipantsLast Reply
Piccolo express
Deborah Martuch, MT(ASCP), CPP
about 2 years ago
30248
Deborah Martuch, MT(ASCP), CPP
about 2 years ago
Quantra
Ashlee Byquist
about 2 years ago
50326
Ashlee Byquist
about 2 years ago
PPM Procedures/Resources
Jennifer Toncray
about 2 years ago
50242
Jennifer Toncray
about 2 years ago