Would anyone like to share their experiences with the Siemen's EPOC for blood gas testing? Pro's and Con's would be greatly appreciated!
76 Replies
Displaying items 61-75 of 76 in total
Reply
Subgroup Membership is required to post Replies
Join POCT Listserv now
Suggested Posts
Topic | Replies | Likes | Views | Participants | Last Reply |
---|---|---|---|---|---|
Piccolo express | 3 | 0 | 262 | ||
Quantra | 5 | 0 | 339 | ||
PPM Procedures/Resources | 5 | 0 | 256 |
Thomas Miller, would you mind to share the IQCP for the EPOC with me as well? Kaylaisaacs@sjhlex.org. I appreciate your willingness to help everyone! Have a wonderful day!
Thomas please send me a copy of your procedure and IQCP. Thank you so mucg for sharing.
dpauff@limamemorial.org
Would somebody help by sharing the IQCP for EPOC. We are also shifting over to Siemens EPOC testing machine. rejiweb@hotmail.com. Thanks. and appreciate your willingness to help.
Looking into EPOC too. Any advice, information, SOP, IQCP would be greatly appreciated. elizabeth.martin2@pennmedicine.upenn.edu
Hi Thomas,
I would love a copy of your IQCP as well.
Thanks for sharing.
nubiawood1@gmail.com
Hi Thomas!
Looks like you are the famous guy on here that knows much about the EPOC. If you don't mind, could you share your SOP and IQCP with me. My email is Elizabeth.Owusu-Ansah@holycrosshealth.org
I appreciate your help.
Hi Thomas,
I am new to the user group and we are also looking into the EPOC. Would you mind sharing your IQCP and SOP for the EPOC.
My email is Debbie-Ann.chin@atlantichealth.org
Hello Thomas,
Could you also provide me with a copy of your IQCP and SOP? It'll be greatly appreciated!
Email: Jason.king@atlantichealth.org
Thomas, my hospital is switching over to the EPOC system. We begin validations next week. Would mind sharing your SOP and IQCP with me as a guide?
jdeal2@capefeavalley.com
Hello Thomas,
We are also moving to the EPOC and any policies/procedures or IQCP would be extremely helpful. If you wouldn't mind sharing. Nina.a.lopez@medstar.net
Thanks in advance.
If I missed anyone please Email me, millerth@einstein.edu
Thomas Miller thanks so much for sharing.
Question can i use istat to correlate with Epoc. since there is issue with the blue card for Chem8 and CG4, can i still uset it for correlation. We are accredit by CLIA, i called them and see waht they can say, CLIA say they cannot advice me about it I am concernned about my correlation because i don't have anything else beside the Istat to correlate. Any advice please!
I would consult with my pathologist on this. We actually correlate with the Lab analyzer, even though it is a bit more strenuous and different sample types. Our thinking is to ensure continuity for result interpretation or to establish any bias that the Dr's should be aware of. For example the lactate's are very sensitive to "tweeks" from siemens. When mmol/l are uses for reporting units small changes can mean a lot at decision levels. This process also gives us flags to send to the vendor (even though the response sucks) when we see a shift and gives us a check on the gen lab as well. Sometimes its not siemens but a lab problem.
We will use the Chem 8 and CG4 to correlate our new devices (Epoc and ChemSTAT). We have years of data on this at many of our sites. For the ED UC, we don't have a lab instrument to compare to, as it is free-standing.