Interfacing Medtronic ACT Plus

0 followers
0 Likes

Hello,


Does anyone have their Medtronic ACT Plus devices interfaced?  I have some questions about the operator workflow that Medtronic seems unable to answer.


Can testing personnel accept or reject a result?


Thanks!


Danielle

1 Reply

Our Medtronic HMS Plus's are interfaced. We have Rals and Epic. I had IT build it where the results stopped in the flagged results in Rals (in case of data entry error by operator). I review and send to Epic. I can also edit in the operators ID once in Epic.


No, testing personnel cannot accept or reject a result.


 


Kenda Webb, MT(ASCP)


Supervisor, POC


Laboratory


_______________________________________________________________


St Francis Hospital


6161 S Yale Ave, Tulsa, OK 74136


Phone: 918-494-2265   |   Fax: 918-494-7238 

Reply
Subgroup Membership is required to post Replies
Join POCT Listserv now
Danielle Ennis
about 5 years ago
1
Reply
0
Likes
0
Followers
302
Views
Liked By:
Suggested Posts
TopicRepliesLikesViewsParticipantsLast Reply
Piccolo express
Deborah Martuch, MT(ASCP), CPP
over 2 years ago
30262
Deborah Martuch, MT(ASCP), CPP
over 2 years ago
Quantra
Ashlee Byquist
over 2 years ago
50339
Ashlee Byquist
over 2 years ago
PPM Procedures/Resources
Jennifer Toncray
over 2 years ago
50256
Jennifer Toncray
over 2 years ago