Manufacture QC Ranges

6 followers
0 Likes

Do you use the manufacture range for the glucometer? or do you adjust it? 


I appreciate any feedback. 

4 Replies

I use the manufacturer's range.

We think the manufacturer's ranges are too wide. We use the TAE, total allowable error from CAP. 10% from the mean or 6 mg/dl. For use it works well, as we can address issue before the values are out of the manufacturer's ranges. We continue to see QC level 1 values continuously below the mean. 

I use manufacturer's range.

We use the manufacturer's range. 

Reply
Subgroup Membership is required to post Replies
Join POCT Listserv now
Ashley Liles
almost 5 years ago
4
Replies
0
Likes
6
Followers
53
Views
Liked By:
Suggested Posts
TopicRepliesLikesViewsParticipantsLast Reply
Piccolo express
Deborah Martuch, MT(ASCP), CPP
over 2 years ago
30262
Deborah Martuch, MT(ASCP), CPP
over 2 years ago
Quantra
Ashlee Byquist
over 2 years ago
50339
Ashlee Byquist
over 2 years ago
PPM Procedures/Resources
Jennifer Toncray
over 2 years ago
50256
Jennifer Toncray
over 2 years ago