Invalid FLU results on Sofia 2
4 followers
0 Likes
Has anyone noticed that when a patient is Covid positive the Sofia 2 FLU A/B will give Invalid results?
We do not use the Combo kit (yet!).
I have a call in to Quidel, but I thought I would ask the Group, too!
We do not use the Combo kit (yet!).
I have a call in to Quidel, but I thought I would ask the Group, too!
4 Replies
Reply
Subgroup Membership is required to post Replies
Join POCT Listserv now
Suggested Posts
Topic | Replies | Likes | Views | Participants | Last Reply |
---|---|---|---|---|---|
Piccolo express | 3 | 0 | 248 | ||
Quantra | 5 | 0 | 326 | ||
PPM Procedures/Resources | 5 | 0 | 242 |
1. Most current firmware and assay software.
2. We had a problem with an analyzer with multiple Invalid results - problem with the optics. If staff are "slamming" the cassette holder door into the analyzer with a full test cartridge splashing can occur onto the optics causing intermittent Invalids.
3. Pipetting was a big issue initially - please see the attachment from Quidel regarding proper pipetting techniques.
Good Luck