Does anyone use ABL 800s in the POC setting?  If so, what AMRs are you using?  

2 Replies

Yes, and we use the AMR set by the most recent run of cal verifications (VK-R7), average of each recovered point.  Also add on a low pO2 cal ver from an alternate vendor, to bring the low end of pO2 to around 10-15 mmHg.

So do you just run the ABL as high complexity since you're out of manufacturers listed AMR?

Reply
Subgroup Membership is required to post Replies